  • 仪器编号
    规格
    生产厂家 Shimadzu 岛津 型号 UV-2700
    制造国家 日本 分类号
    放置地点 4号楼220房间 出厂日期 2019/01
    购置日期 2019/01 入网时间 2019/06/25
    主要规格及技术指标
    测试波长范围: 185-900nm(更换PMT,可延伸至1150nm)
    衍射光栅刻线数: 1300 lines/mm
    波长准确性: ±0.1nm(656.1nm)±0.3nm(全波段)
    波长重复精度: ±0.05nm
    波长扫描速度: 波长移动速度: 14000nm/min; 最大扫描速度:4500nm/min;
    波长设定: 扫描开始波长和扫描结束能够以1nm单位设置;其它为0.1nm单位
    光源切换波长: 和波长同步自动切换290.0 nm~370.0 nm
    谱带宽度: 0.1/ 0.2/ 0.5/ 1/ 2/ 5nm L2/L5(低杂散光模式)
    分辨率: 0.1nm
    杂散光: KCI < 1%T  (198nm)
    NaI < 0.00005%T (220nm)
    NaNO2 < 0.00002%T (340nm)
    测光方式: 双光束测光方式
    测光类型: 吸光度(Abs),透射率(%),反射率,能量(E)
    测光范围: 吸光度:-8.5~8.5 Abs
    光度准确性 ±0.002Abs(0.5Abs)
    ±0.003Abs(1Abs)
    ±0.006Abs(2.0Abs)
    ±0.3%T
    光度重现性 ±0.001Abs(0.5Abs)
    ±0.001Abs(1Abs)
    ±0.003Abs(2Abs)
    ±0.1%T
    噪音 0.00005Abs RMS (500nm)
    基线稳定性 < 0.0003Abs/hour
    基线平直度 ±0.0004Abs(200-860nm)
    记录范围: 吸光度-9.999~9.999 Abs; 透射率-999.9~9.999%
    漂移:小于0.0003Abs/h
    基线校正:计算机自动校正(电源启动时,自动存储备份的基线,可以再校正)

    主要功能及特色 光谱测量,光度测量,测试范围185-900nm,可用于液体样品的测试。
    主要附件及配置 光源: 50W卤素灯和氘灯(插座型)
    检测器:光电倍增管


  • 技术资料:https://iscps.westlake.edu.cn/info/1131/1301.htm

    样品要求:液体,玻璃,薄膜等能透光样品

  • 类型

    项目

    细则

    校内

    校外

    说明

    自主上机

    自主上机机时费

    液体、常温全波段测试

    40/

    80/

    /

    培训

    仪器培训

    一对一/一对二培训;同等内容的第二次培训,收费1.5

    80/

    120/

    /

    委托测试 光谱/光度测试 液体测试 80/ 120/ /

     

  • Instrument No. LHA19040018                                                             Specification

    Manufacture    Shimadzu                                                                     Model                        UV-2700

    Made in           Japan                                                                           Classification No 

    Location          4#220                                                                          Manufacture date

    Purchase Date 5/03/2019                                                                     Internet access time   25/06/2019

    Main Technical specifications

    Wavelength range: 185~900 nm

    Diffraction grating number of lines: 1300 grooves/mm

    Wavelength accuracy: ±0.1nm656.1nm ±0.3nm (All wavelength)

    Wavelength repeatability: ±0.05nm

    Wavelength scanning speed: 1. Wavelength moving speed : 14000nm/min  2. Wavelength scanning speed: 4500nm/min

    Wavelength setting: the wavelength at the beginning of scan and t at the end of scan can be set at 1nm; Other are 0.1nm

    Light switching wavelength: synchronization automatically switch 290.0 nm~370.0 nm

    Band width: UVVIS0.1/ 0.2/ 0.5/ 1/ 2/ 5 nm L2/L5 8 (Low stray light mode)

    Resolution: 0.1nm

    Stray light: KCl<1%T (198nm )

    NaI<0.00005%T (220nm)

    NaNO2<0.00002% (340nm,)

    Measurement mode: Dual beam system

    Measurement type: AbsorbanceAbs),T(%),ReflectivityEnergyE

    Absorbance Range :-8.5~8.5 Abs

    Photometric accuracy±0.002Abs (0.5Abs)

    ±0.003Abs (1Abs)

    ±0.006Abs (2.0Abs)

    ±0.3%T (0.5Abs)

    Photometric repeatability±0.001Abs.( 0.5Abs)

    ±0.001Abs (1Abs)

    ±0.003Abs (2.0Abs)

    ±0.1%T (0.5Abs)

    Noise 0.00005Abs RMS (500nm)

    Baseline stability < 0.0003Abs/hour

    Baseline flatness ±0.0004Abs200-860nm

    Recording range: absorbance -9.999~9.999 Abs; Transmission 999.9 ~ 9.999%

    Driftingless than 0.0003Abs/h

    Baseline correction: automatic correction by computer

     

    Main functions and features

    Spectral measurement, photometric measurement. Range: 185-900nm. Sample: liquid, glass, film sample test.

    Accessories and configuration

    Light source: 50W halogen lamp and deuterium lamp (socket type)

    Detector: photomultiplier

     

  • Information: https://iscps.westlake.edu.cn/info/1131/1301.htm

    Sample requirements: liquid, glass, thin film, etc

     


  • Operation Types

    Items

    Requirements

    Internal prices

    External prices

    Comments

    Independent   operation

    Liquid, film RT- measurement in   Full wavelength

    After   being well-trained and authorized, Make an appointment online

    ¥40/hour

    ¥80/hour

    /

    Training

    Instrument   training

    One-on-one   or one-on-two training. The first training session is normal charge. Any   subsequent training for the same instrument will be charged as a rate of 1.5 times.  

    ¥80/hour  

    ¥120/hour

    /